Equipment Storage And Maintenance; Disinfectant Information - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Equipment Storage and Maintenance

Follow the storage and maintenance procedures for dialysis equipment used for Intermittent
Hemodialysis (IHD) in the ICU setting.
Warning! Possible explosion hazard if used in the presence of flammable anesthetics.
Storage Location
If the dialysis unit is incorporated in the hospital, the equipment should be stored so that it
will not be damaged. The maintenance of the equipment is normally part of the duty of the
dialysis service technician. Depending on the disinfectant and the storage time, frequent
flushing of the equipment is necessary.
If it is more convenient to store the equipment close to the ICU, the dedicated storage space
should have an access to tap water, power and a drain. The room should be well vented, if
disinfectant is used.
Storage Preparation
Before storing the 2008T hemodialysis machine, the hydraulics should be disinfected. It is
also necessary to wipe the external parts of the machine with a surface cleaner. Frequency of
disinfection and length of the dwell time depend on the disinfectant and shall be determined
by acceptable culture result. The following table lists commonly used procedures to store
equipment and then return it from storage. Validate your own procedure according to the
hospital policy. After prolonged storage, use a bleach disinfectant prior to patient use.
Table 40 – Disinfectant Information
Disinfectant
Formaldehyde
Bleach
Puristeril 340
Renalin
Heat
2008T Machine Operator's Manual P/N 490122 Rev N
Program
Dwell time
Fill Hydraulic
Unlimited
(Service Mode)
Chemical Disinfection
Only for time of
program
disinfection program
Chemical Disinfection
Only for time of
program
disinfection program
Heat Disinfection
Recirculate and shut
program
machine off
Appendix C
Retest and Repeat
3 – 4 weeks
24 hours
24 hours
24 hours
263