Wd: 24V Rcvr Err Long; Wd: 24V Rcvr Err Short; Wd: Fail Long Pulse; Wd: Fail Short Pulse - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Message
Purpose of Message
Wait, OLC Aborting
Changed blood pump flow rate,
changed dialysate flow rate, or
unstable conductivity caused on line
clearance test to stop.
Wait: Rinsing Line
The machine is rinsing the
concentrate lines prior to a cleaning
or disinfecting program.

WD: 24v Rcvr Err Long

Electronic self-test failure.

WD: 24v Rcvr Err Short

Electronic self-test failure.

WD: Fail Long Pulse

Electronic self-test failure.

WD: Fail Short Pulse

Electronic self-test failure.
217
Troubleshooting
Type
Action Required
Advisory
Wait until stable conditions for OLC test to begin again
Advisory
Advisory only. No action is required. Line rinsing takes about 45
seconds.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
ALARMS
TROUBLE-
SHOOTING
2008T Machine Operator's Manual P/N 490122 Rev N