Fresenius Medical Care 2008T Operator's Manual page 208

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

208
Troubleshooting
SN BP Comm. Timeout
Time out when receiving Intel-Hex-
line or overflowed received buffer
SN BP Del. Rate Alarm
Actual speed-read back analog
voltage at X348/10 is out of limits
SN BP Direction Alarm
Pump is turning in the wrong direction
SN BP EEPROM Error
EEPROM error
SN BP EPROM Error
EPROM CRC error
Error erasing Flash ROM while in
SN BP Erasing Error
Service Mode
SN BP Flash Error
Error copying data into Flash ROM
while in Service Mode
SN BP Level Up Alarm
Pressure increase when the Level Up
key is pressed
SN BP Pressure Alarm
Arterial pressure-read back analog
voltage at X348/7 is out of limits
SN pressure-read back analog
voltage is out of limits
SN BP RAM Error
RAM check error
SN BP Rate Alarm
Pump is not reaching speed at
maximum voltage
SN BP Receiving Alarm
Error in received Intel-Hex-line
SN BP ROM Error
Flash ROM CRC error
SN BP Rotation Error
Pump rotor turning when it should not
be for a second time
SN BP Set Rate Alarm
Set speed-read back analog voltage
at X348/14 is out of limits
SN BP Stop Alarm
Pump rotor turning when it should not
be
SN BP Tach Alarm
Optical tachometer not in range
SN BP Task Error
Software task was not completed
correctly
SN BP Timer Error
50 ms second time period exceeded
SN BP Update Error
Transmit error during Flash update
while in Service Mode
2008T Machine Operator's Manual P/N 490122 Rev N
Alarm
See message A.27
Alarm
See message A.21
Alarm
See message A.13
Alarm
See message E.05
Alarm
See message E.01
Alarm
See message E.98
Alarm
See message E.97
Alarm
See message A.25
Alarm
See message A.22
Alarm
See message E.03
Alarm
See message A.11
Alarm
See message A.28
Alarm
See message E.02
Alarm
See message E.23
Alarm
See message A.20
Alarm
See message A.29
Alarm
See message A.24
Alarm
See message E.15
Alarm
See message E.14
Alarm
See message E.99