Chapter 3—Setting Treatment Parameters
actual conductivity of the dialysate is displayed on the right side, above the Conductivity bar
graph.
Most dialysate or dialysate-related alarm parameters are accessed from the "Dialysate"
screen. Unless otherwise described, enter or change a dialysate-related value by following
the procedure described in "Entering a Parameter" on page 62.
What to do from this screen...
Enter the prescribed dialysate settings for:
Concentrate type
Base Na
Bicarbonate level
Sodium Variation (SVS) profile
Set Alarm limits for:
Low Acid/Bicarbonate alert
Position and width of Conductivity Alarm window
Turn Dialysate Flow on or off (software version 2.34 or later)
Warning! The specific concentrate and sodium and bicarbonate settings must be prescribed
by a physician.
Note: If the machine is set up for use with Citrasate, a 'Citrate' meter box will be displayed
in the dialysate constituent list.
Acid Concentrate Type
Theoretical
Conductivity (TCD)
of the Final
Dialysate at the
+
displayed Base Na
and Bicarbonate
values.
Electrolyte Profile of
Selected
Concentrate at the
base sodium and
bicarbonate levels
Figure 29 – The Dialysate Screen
64
+
level
2008T Machine Operator's Manual P/N 490122 Rev N
Alarm Window
Conductivity
Scale
Actual Conductivity