Base Na+ Greater Than Max. Value; Bibag: +5 V Error/Bibag: -5 V Error/Bibag: +12 V Error; Bibag: Bag Leak; Bibag: Bag On - Fresenius Medical Care 2008T Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008T:
Table of Contents

Advertisement

Message
Purpose of Message
Base Na+ greater than
Entered Base Na+ parameter is
max. value
higher than allowed.
Base Na+ has been set
The operator has attempted to set a
to min.
Base Na+ lower than allowed.
bibag: +5 V Error
Electronic self-test, power supply
limits exceeded.
bibag: -5 V Error
Electronic self-test, power supply
limits exceeded.
bibag: +12 V Error
Electronic self-test, power supply
limits exceeded.

bibag: Bag Leak

A leak has been detected in the bag.

bibag: Bag On

The bag is on the connector when
user is either attempting to run a
cleansing/ disinfecting program or
using acetate.
Or
The blue bicarbonate connector is out
of the bicarbonate port when a bag is
on the bibag connector.
bibag: Bic Pump Locked
The bicarbonate pump has been air
locked for over two minutes.
bibag Board Failure
bibag Interface Board cannot boot up.
165
Troubleshooting
Type
Action Required
Dialog
The Base Na+ will be set to the highest allowed Na+ level. Press
Message
CONFIRM to clear message and accept the maximum allowed value.
Verify that the value is acceptable or enter new value.
Dialog
The Base Na+ will be set to the lowest allowed Na+ level. Confirm if
Message
value is acceptable or enter new value.
Alarm
Turn machine power off and back on. If the alarm is not cleared, take
the machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If the alarm is not cleared, take
the machine out of service and alert a qualified service technician.
Alarm
Turn machine power off and back on. If the alarm is not cleared, take
the machine out of service and alert a qualified service technician.
Water
Open bibag door, wait 30 seconds to relieve the pressure, and
Alarm
remove the bag. Place a new bag on the connector and close the
bibag door.
Water
Remove the bag from the bibag connector if using acetate or
Alarm
attempting to run a cleansing/disinfecting program.
Or
Plug the blue bicarbonate connector back into the bicarbonate port on
the machine.
Water
If during treatment, rinse back the patient's blood and disconnect the
Alarm
patient from the machine. Run a Rinse program to clear the alarm. If
the alarm is not cleared, take the machine out of service and alert a
qualified service technician.
Alarm
Turn machine power off and back on. If alarm is not cleared, take the
machine out of service and alert a qualified service technician.
ALARMS
TROUBLE-
SHOOTING
2008T Machine Operator's Manual P/N 490122 Rev N