Machine Specifications - Fresenius Medical Care 2008K Operator's Manual

Hemodialysis machine
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Machine Specifications

Dimensions
Floor space
Height
Total weight
Operating conditions
Storage conditions
Electrical
Power Supply—Main
Power Consumption
Fuses
External Connections
Heat Dissipation to Room
Electromagnetic compatibility
Electrical safety (UL 60601-1)
Protection against electric shock
Leakage currents
Water
Back Flow Prevention
Water Pressure
Water Temperature
2008K Hemodialysis Machine Operator's Manual Rev P
Approximately 54 cm wide by 63 cm deep
133 cm
Approximately 73 kg
60 – 100 °F (15.5 – 38 °C)
Room temp, 6 months, Do not freeze
Single phase AC 117 V ±10% 60 Hz ±3 Hz must be
connected to a circuit which is equipped with a hospital
grade receptacle and is protected by circuit breaker and
ground fault interrupter (GFI). Resistance from chassis to
ground must be < 0.2 ohm.
Does not exceed 12.5 amps
6.3 amp medium blow fuse, 2 each
16 amp double pole rocker switch circuit breaker for
heater
A.
Alarm input
B.
External alarm light or traffic light beacon
C.
Isolated RS232 and leakage current isolation per
UL 60601-1 between the machine and external
computer
600 to 700 BTU/hr
See EMC Declaration on page 224
Type: Safety class I Degree: Type B
Type CF: Only BPM Blood Pressure Cuff
According to UL 60601-1
Integral back flow prevention provided by external vent to
atmosphere in water inlet circuit.
Min 20 psi; max 105 psi
Min 10 °C; max 25 °C
Machine Specifications