Basic Operations for RF Devices Measurement
STEP 4. Connecting Test Fixture
STEP 4. Connecting Test Fixture
Connect the test fixture to the 7-mm terminal of the test head by following these steps. In
this section, the method of connecting the 16197A test fixture is described as an example.
When using other test fixtures, refer to the Operation Manual of the test fixture.
Step 1. As shown in Figure 3-42, turn the 7-mm connector nut of the test head counterclockwise
until the connector sleeve is fully retracted.
Figure 3-42
Retracting 7-mm connector sleeve
Step 2. Set the two mount posts of the test head to the two holes of the test fixture and set the
7-mm terminal of the test head to 7-mm connector of the test fixture (Figure 3-43: 2).
Step 3. Turn the 7-mm connector nut of the test head counterclockwise to connect it securely
(Figure 3-43: 3). Use both hands to turn the connector nut because the space between the
test head and the test fixture is narrow.
Figure 3-43
Connecting test fixture (16197A)
70
Chapter 3