2.1
Figure 2-1
The same setup can be
used to measure the
Insertion Loss.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Taking Polarization Dependent Measurements
Measuring Polarization Dependent Loss
Measuring Polarization Dependent Loss
This section covers the use of the applications "Pol. Depend. Loss"
and "PDL / Insert. Loss". You can measure how the optical loss of
passive components is affected by different states of polarization.
PDL Measurement Setup
Signal Processing and Display
Head Input A
Head Input A
Head Input B
Head Input B
Setup for PDL Measurements
1 Connect the optical output to the input of the device under test
(DUT).
2 Connect the optical head to the output of the DUT.
41
Coupler
Internal
Source(s)
Pol. Ctrl.
3 dB
Optical Input
Optical Input
Optical Output
Optical Output
Device
under
Test