Procedure for Using Front Panel Keys

Measuring OPEN Compensation Data (16197A)

Perform OPEN compensation to correct stray admittance due to the test fixture.
Step 1. Set the DUT connection terminal of the test fixture to the OPEN state by following the
procedure described in Figure 3-47. When using other test fixtures, follow the procedure
described in the Operation Manual of the test fixtures.
Step 2. Press the
Step 3. Press the
Comp Menu
Step 4. Press the
Step 5. A √ mark appears on the left side of the
compensation data measurement.
NOTE
The measurement of OPEN and SHORT compensation data can be stopped by pressing the
the

Measuring SHORT Compensation Data (16197A)

Perform SHORT compensation to correct residual impedance due to the test fixture.
Step 1. Connect a short device to the test fixture. Follow the procedure illustrated in Figure 3-50 to
set the DUT connection terminal of the test fixture to the SHORT state. When using other
test fixtures, follow the procedure described in the Operation Manual of the test fixture.
Step 2. Press the
the
Step 3. A √ mark appears on the left side of the
SHORT compensation data measurement.
Chapter 3
key in the STIMULUS block.
[Cal/Compen]
or
key in the ENTRY/NAVIGATION block to move the cursor to
button and then press the
or
key to move the cursor to the
key (Figure 3-49).
or
key to move the cursor to the
key during measurement.
or
key to move the cursor to the
key (Figure 3-51).
Basic Operations for RF Devices Measurement
STEP 6. Fixture Compensation
key (Figure 3-48).
button and press the
Meas Open
button upon completion of the OPEN
Meas Open
button and then pressing
Abort Comp Meas
button and then press
Meas Short
button upon completion of the
Meas Short
79