Basic Operations for RF Devices Measurement
STEP 3. Calibration

Measuring SHORT Calibration Data

Use the 0 Ω (SHORT) standard to perform SHORT calibration by following these steps.
Step 1. Turn the 0 Ω (SHORT) standard clockwise with the provided torque wrench to connect it
securely to the 7-mm terminal (Figure 3-33).
Step 2. Press the
or
the
key (Figure 3-34).
Step 3. A √ mark appears on the left side of the
SHORT calibration data measurement.
Step 4. Turn the 0 Ω (SHORT) standard counterclockwise to remove it.

Measuring LOAD Calibration Data

Use the 50 Ω (LOAD) standard to perform LOAD calibration by following these steps.
Step 1. Turn the outside connector nut of the 50 Ω (LOAD) standard counterclockwise to fully
retract the inside connector sleeve (Figure 3-35).
Step 2. Turn the 50 Ω (LOAD) standard clockwise with the provided torque wrench to connect it
securely to the 7-mm terminal (Figure 3-36).
Step 3. Press the
or
the
key (Figure 3-37).
Step 4. A √ mark appears on the left side of the
calibration data measurement.
Step 5. Turn the 50 Ω (LOAD) standard counterclockwise to remove it.

Measuring LOW-LOSS CAPACITOR Calibration data

The LOW-LOSS CAPACITOR calibration should be performed for high Q (or low D:
dissipation factor) measurements at high frequencies. The LOW-LOSS CAPACITOR
calibration allows high accuracy for phase measurements. This calibration can be skipped
if you do not need it for your purposes.
Step 1. Turn the LOW-LOSS CAPACITOR clockwise with the provided torque wrench to connect
it securely to the 7-mm terminal (Figure 3-38).
Step 2. Press the
or
press the
key (Figure 3-39).
Step 3. A √ mark appears on the left side of the
LOW-LOSS CAPACITOR calibration data measurement.
Step 4. Turn the LOW-LOSS CAPACITOR counterclockwise to remove it.
68
key to move the cursor to the
Meas Short
key to move the cursor to the
Meas Load
key to move the cursor to the
Meas Low Loss C
button and then press
Meas Short
button upon completion of the
button and then press
Meas Load
button upon completion of the LOAD
button and then
Meas Low Loss C
button upon completion of the
Chapter 3