Measuring Short Calibration Data - Agilent Technologies E4991A Installation And Quick Start Manual

Impedance/material analyzer
Hide thumbs Also See for E4991A:
Table of Contents

Advertisement

Step 4. Turn the 0 S (OPEN) standard counterclockwise to remove it.
NOTE
Clicking the
LOW-LOSS CAPACITOR (optional) calibration data stops the measurement.

Measuring SHORT Calibration Data

Use the 0 Ω (SHORT) standard to perform SHORT calibration by following these steps.
Step 1. As shown in Figure 3-33, turn the 0 Ω (SHORT) standard clockwise with the provided
torque wrench to connect it securely to the 7-mm terminal.
Connecting 0 Ω (SHORT) standard
Figure 3-33
Step 2. Click the
Figure 3-34
Step 2
Step 3. A √ mark appears on the left side of the
SHORT calibration data measurement.
Step 4. Turn the 0 Ω (SHORT) standard counterclockwise to remove it.
Chapter 3
Basic Operations for RF Devices Measurement
button during measurement of OPEN, SHORT, LOAD, and
Abort Cal Meas
button (Figure 3-34).
Meas Short
STEP 3. Calibration
button upon completion of the
Meas Short
63