Measuring Insulation Resistance Of Capacitor; Measurement Setup; Measurement Procedure - Agilent Technologies 4339B User Manual

High resistance meter
Hide thumbs Also See for 4339B:
Table of Contents

Advertisement

Measuring Insulation Resistance of Capacitor

This example shows the procedure to measure insulation resistance of capacitor after charged 1
minute. Using the test sequence measurement function reduces the measurement complexity.
DUT
Chip ceramic capacitor

Measurement Setup

Measurement Parameter : R
Measurement Range : Auto range mode
Test Voltage : 100 V
Use the measurement sequence single mode (measure after charged for 1 minute.)

Measurement Procedure

1. Reset the 4339B.
a. Press
b. Select
using
Yes
2. Connect the test xture to the UNKNOWN terminals as follows:
* For detail, see \16339A Component Test Fixture" in Chapter 2.
3. Perform calibration.
Press
3-2
Measurement Examples
.
or
and press
.
Requirements
Test Fixture : 16339A, SMD module
.
4339B