Agilent Technologies E4991A Installation And Quick Start Manual page 116

Impedance/material analyzer
Hide thumbs Also See for E4991A:
Table of Contents

Advertisement

Basic Operations for Dielectric Measurement
STEP 6. Calibration
Step 5. Set the MUT connection plane of the test fixture to the SHORT state (Figure 4-6).
Figure 4-6
SHORT state (16453A)
Step 6. Click the
of the SHORT calibration measurement, a √ mark appears on the left side of the
Short
NOTE
During calibration data measurement, the message "Wait-Measuring Cal Standard"
appears at the left end of the status bar at the bottom of the screen.
116
button to start measuring SHORT calibration data. Upon completion
Meas Short
button.
Meas
Chapter 4