Basic Operations for Dielectric Measurement
STEP 6. Calibration
Step 5. Set the MUT connection plane of the test fixture to the SHORT state (Figure 4-6).
Figure 4-6
SHORT state (16453A)
Step 6. Click the
of the SHORT calibration measurement, a √ mark appears on the left side of the
Short
NOTE
During calibration data measurement, the message "Wait-Measuring Cal Standard"
appears at the left end of the status bar at the bottom of the screen.
116
button to start measuring SHORT calibration data. Upon completion
Meas Short
button.
Meas