STEP 4. Connecting 16453A

Connect the 16453A test fixture to the 7-mm terminal of the test head by following these
steps.
Step 1. As shown in Figure 3-42 on page 70, turn the 7-mm connector nut of the test head
counterclockwise until the connector sleeve is fully retracted.
Step 2. Tighten the two small screws of the fixture holder to secure the fixture holder to the test
fixture body (Figure 4-4, 2).
Step 3. Connect the 7-mm connector of the test fixture to the 7-mm terminal of the test head
(Figure 4-4, 3).
Step 4. Tighten the two large screws of the fixture holder to secure the test fixture to the test head
(Figure 4-4, 4).
Figure 4-4
Connecting the 16453A