Basic Operations for Dielectric Measurement
STEP 6. Calibration
Step 9. Connect the load standard supplied with the 16453A test fixture to the test fixture by
inserting it between the electrodes of the test fixture (Figure 4-8).
Figure 4-8
Connecting LOAD standard (16453A)
NOTE
When connecting a load standard or a MUT to the test fixture, make sure that it only comes
into contact with the test fixture's electrodes. Also, be careful not to give the upper
electrode horizontal pressure by moving the load standard or the MUT while it is in
position between the electrodes.
118
Chapter 4