Step 7. Set the MUT connection plane of the test fixture to the OPEN state (Figure 4-7).
Figure 4-7
OPEN state (16453A)
Step 8. Click the
of the OPEN calibration data measurement, a √ mark appears on the left side of the
Open
Chapter 4
button to start measuring OPEN calibration data. Upon completion
Meas Open
button.
Basic Operations for Dielectric Measurement
STEP 6. Calibration
Meas
117