Step 8. Measuring Dut And Analyzing Measurement Results; Applying Dc Bias (Option 001) - Agilent Technologies E4991A Installation And Quick Start Manual

Impedance/material analyzer
Hide thumbs Also See for E4991A:
Table of Contents

Advertisement

Basic Operations for RF Devices Measurement

STEP 8. Measuring DUT and Analyzing Measurement Results

STEP 8. Measuring DUT and Analyzing Measurement
Results
After performing calibration and compensation, the measurement results are displayed on
the screen when the DUT is connected to the test fixture. If you obtain the correct trace,
you can analyze the measurement results by using the marker and the equivalent circuit
analysis function. This section describes the procedure by using only the mouse and
keyboard.

Applying dc Bias (Option 001)

When selecting "(4) Dc Bias (Current) Characteristics of Ls-Q (Option 001)" on page 58 in
"STEP 2. Setting Measurement Conditions", follow these steps to apply dc bias. For
settings of the dc bias source level and dc bias limit, refer to Chapter 3, "Setting
Measurement Condition" in the E4991A Operation Manual.
WARNING
Never touch the DUT or the electrodes of the test fixture while dc bias is applied.
Step 1. Click
Step 2. Click the
bias ON places the measurement in the Hold Mode (mode that does not accept triggers).
Figure 3-58
Steps 2, 3
Step 3. Confirm that the display of the status bar on the bottom of the screen changes to "Bias ON"
and the trigger system is in the Hold Mode (Figure 3-58).
Step 4. Set the trigger to start the measurement.
o Perform the single measurement.
82
on the
Source...
Stimulus
button to change the button's display to
Dc Bias
1. Click
on the
Single
Trigger
menu (Figure 3-15).
[On]
menu.
(Figure 3-58). Making dc