Basic Operations for Magnetic Measurement
STEP 10. Changing Sweep Conditions
STEP 10. Changing Sweep Conditions
When the measurement points for calibration/fixture compensation are user-defined
frequency/user-defined power points, start measurement with "STEP 4. Calibration" on
page 133 after changing the sweep conditions. When the measurement points for
calibration/fixture compensation are fixed frequency/fixed power points, start
measurement with "STEP 9. Measuring MUT and Analyzing Measurement Results" on
page 140 after changing the sweep conditions.
NOTE
When the measurement points for calibration/fixture compensation are fixed
frequency/user-defined power points, start measurement with "STEP 4. Calibration" on
page 133 after changing the oscillator level. When you change other sweep conditions,
start measurement with "STEP 9. Measuring MUT and Analyzing Measurement Results"
on page 140.
STEP 11. Measuring Other MUTs
If you measure another MUT of the same size as the one used in the previous
measurement, start measurement with "STEP 8. Mounting MUT" on page 139. If you use
the same test fixture to measure a MUT of a different size, start with "STEP 7. Entering
MUT Dimensions" on page 138. When using a different test fixture, start with "STEP 5.
Connecting 16454A" on page 135.