STEP 5. Connecting 16454A

The 16454A test fixture has two sizes: Small and Large. In addition, it has four MUT
holders. Select the size of the MUT holder that best suits your needs (see Table 5-4 and
Figure 5-5). This selection determines whether the 16454A (Small) or 16454A (Large) test
fixture is the appropriate test fixture, as shown in Figure 5-5.
Table 5-4
MUT size for test fixtures
Test fixture
MUT holder
MUT inner diameter
(mm): b
MUT outer diameter
(mm): c
MUT height (mm): h
Figure 5-5
MUT holder sizes
Connect the selected test fixture to the 7-mm terminal of the test head by following these
steps.
Step 1. As shown in Figure 3-42 on page 70, turn the 7-mm connector nut of the test head
counterclockwise until the connector sleeve is fully retracted.
Step 2. Tighten the two small screws of the fixture holder to secure the fixture holder to the test
fixture body (Figure 5-6: 2).
Step 3. Tighten the two large screws of the test holder to secure the test fixture to the test head
(Figure 5-6, 3).
Step 4. Connect the 7-mm connector of the test fixture to the 7-mm terminal of the test head
(Figure 5-6, 4).
Chapter 5
16454A (Small)
A
B
≥ φ 3.1 mm
≥ φ 3.1 mm
≤ φ 8.0 mm
≤ φ 6.0 mm
≤ 3.0 mm
≤ 3.0 mm
Basic Operations for Magnetic Measurement

STEP 5. Connecting 16454A

16454A (Large)
C
≥ φ 6.0 mm
≤ φ 20.0 mm
≤ 8.5 mm
D
≥ φ 5.0 mm
≤ φ 20.0 mm
≤ 8.5 mm
135