Basic Operations for Magnetic Measurement

STEP 7. Entering MUT Dimensions

Step 6. Verify that the display below the
the bottom of the screen changes to "Comp ON".

STEP 7. Entering MUT Dimensions

You must enter the MUT dimensions before you can perform measurement. Use a
micrometer or calipers to measure the outer (see c of Figure 5-2 on page 128) and inner
(see b of Figure 5-2 on page 128) diameters and height (see h of Figure 5-2 on page 128).
NOTE
The 16454A test fixture imposes restrictions on the outer and inner diameters and height of
the MUT (see Table 5-1 on page 128).
Step 1. Click
Step 2. Click the
Step 3. In the
MUT is 3 mm in height, type
Step 4. In the
example, if the inner diameter is 4.5 mm, type
Step 5. In the
example, if the outer diameter is 7-mm, type
NOTE
The selection of a smaller test fixture imposes restrictions on the MUT dimensions that can
be entered.
138
on the
menu.
Utility...
Utility
Material Option Menu
box, enter the height of the MUT (see h in Figure 5-2). For example, if the
Height
[3] [m] [Enter]
box, enter the inner diameter of the MUT (see b in Figure 5-2). For
Inner Diameter
box, enter the outer diameter of the MUT (see c in Figure 5-2). For
Outer Diameter
button changes to
Comp Menu
button.
with the keyboard.
[4] [.] [5] [m] [Enter]
[7] [m] [Enter]
and the status bar at
[ON]
with the keyboard.
with the keyboard.
Chapter 5