Procedure for Using Front Panel Keys
Preparing for Calibration
Prepare for calibration by following these steps.
Step 1. Press the
Step 2. Press the
the
Cal Menu
Step 3. Press the
Freq&Pwr
Step 4. Turn the 7-mm connector nut of the test head clockwise until the connector sleeve is fully
extended (Figure 3-30).
Measuring OPEN Calibration Data
Use the 0 S (OPEN) standard to perform OPEN calibration by following the procedure
described below.
Step 1. Turn the 0 S (OPEN) standard clockwise with the provided torque wrench to connect it
securely to the 7-mm terminal (Figure 3-31).
Step 2. Press the
Step 3. A √ mark appears on the left side of the
calibration data measurement.
Step 4. Turn the 0 S (OPEN) standard counterclockwise to remove it.
NOTE
The measurement of OPEN, SHORT, LOAD, or LOW-LOSS CAPACITOR (optional)
calibration data can be stopped by pressing the
the
Abort Cal Meas
Chapter 3
key in the STIMULUS block.
[Cal/Compen]
or
key in the ENTRY/NAVIGATION block to move the cursor to
button and then press the
or
key to move the cursor to the
key to open the list. Next press the
and press the
key (Figure 3-29).
or
key to move the cursor to the
key (Figure 3-32).
button and the pressing the
Basic Operations for RF Devices Measurement
key (Figure 3-28).
box and then press the
Cal Type
or
key to move the cursor to
button and then press the
Meas Open
button upon completion of the OPEN
Meas Open
or
key to move the cursor to
key during measurement.
STEP 3. Calibration
Fixed
67