Step 2. Verify that the display below the
the status bar on the bottom of the screen changes to "Comp ON" (Figure 3-53).
Figure 3-53
Display of status bar when completing fixture compensation

Checking SHORT Compensation Data

After completing the fixture compensation, use the marker function to check that the
SHORT compensation data have been measured correctly.
Step 1. Specify the trace of the measurement parameter |Z| (in this example it is Trace 1) as the
active trace (* mark) as shown in Figure 3-9 and Figure 3-10. When the parameter |Z| does
not exist in the measurement parameter as in measurement conditions (3) and (4), perform
the check after converting the measurement parameter of Trace 1 to the parameter |Z|. After
the check, return it to the previous measurement parameter.
Step 2. Click
Figure 3-54
Step 2
Chapter 3
Comp Menu
on the
menu (Figure 3-54).
Autoscale
Scale
Basic Operations for RF Devices Measurement
STEP 6. Fixture Compensation
button changes to
[ON]
and the display of
77