Step 1. Preparation For Measurement; Selection Of Mut And Test Fixture; Required Equipment - Agilent Technologies E4991A Installation And Quick Start Manual

Impedance/material analyzer
Hide thumbs Also See for E4991A:
Table of Contents

Advertisement

Basic Operations for Magnetic Measurement

STEP 1. Preparation for Measurement

STEP 1. Preparation for Measurement

Selection of MUT and Test Fixture

With the E4991A, the 16454A test fixture can be used to measure magnetic materials. The
applicable dielectric materials are toroidal cores with a donut shape, such as ferrite
magnets (Figure 5-2).
Figure 5-2
Applicable magnetic materials
Table 5-1
Features of 16454A test fixture
Frequency range
1 M to 1 GHz

Required Equipment

The following equipment is required to perform magnetic measurement.
128
Maximum dc bias current
value
± 50 mA
Magnetic material size
b ≥ φ 3.1 mm
c ≤ φ 20.0 mm
h ≤ 8.5 mm
Chapter 5