Basic Operations for RF Devices Measurement
STEP 2. Setting Measurement Conditions

(1) Frequency Characteristics of |Z|-Ls-Q

You first have to change the measurement conditions from the initial state of the E4991A
as shown in Table 3-2.
Table 3-2

Procedure for Using Mouse and Keyboard

Set up the E4991A by following the procedure given below.
Presetting the E4991A
Step 1. Click
Figure 3-5
Presetting E4991A
46
Setup example for this measurement
Parameter setting
Trace 1
Measurement
Trace 2
parameters
Trace 3
Trace 1
Display formats
Trace 2
Trace 3
Measurement points
Sweep parameter
Sweep type
Source mode
Oscillator level
Sweep range (frequency)
on the
menu to set the initial state (Figure 3-5).
Preset
System
Setup example
|Z|
Ls
Q
Log
Linear
Linear
201 points
Frequency
Log
Current
1 mA
1 MHz to 3 GHz
Initial state
|Z|
θz
Q
Linear
Linear
Linear
201 points
Frequency
Linear
Voltage
100 mV (2 mA)
1 MHz to 3 GHz
Chapter 3