Basic Operations for RF Devices Measurement
STEP 8. Measuring DUT and Analyzing Measurement Results

Detecting Maximum Value

This section describes how to search for the maximum value (self-resonance point) of the
measurement parameter |Z| by using the setup example of "(1) Frequency Characteristics
of |Z|-Ls-Q" on page 46.
Step 1. Specify the trace (measurement parameter) to be read as the active (refer to Figure 3-9,
Figure 3-10). In this case, make Trace 1 (the measurement parameter |Z|) active.
Step 2. Click
Figure 3-72
Step 2
Step 3. Select
Figure 3-73
Step 3
Table 3-11
Search Type
Search Type
Maximum
Minimum
Target
Positive Peak
Negative Peak
Step 4. Click the
Step 5. Active marker moves to the maximum value on the active trace (Figure 3-74).
Step 6. Read the marker value displayed in upper right area of the screen.
92
on the
Function...
Marker
in the
Maximum
Search Type
box settings and search points
box
Search point
Point where the trace value on the active trace is maximum
Point where the trace value on the active trace is minimum
Point where the trace value on the active trace is set to a target
Point where the trace value on the active trace is a positive
peak
Point where the trace value on the active trace is a negative
peak
button (Figure 3-73).
Search
menu (Figure 3-72).
box
(Figure
3-73).
Chapter 3