STEP 4. Calibration

Calibration is performed by using the 7-mm terminal of the test head as the calibration
reference plane (Figure 5-4). Unlike impedance measurement, fixture compensation after
calibration requires only SHORT compensation (see Table 5-3).
Figure 5-4
Error model of 16454A test fixture
Table 5-3
Differences in calibration and fixture compensation between impedance and
magnetic measurement
Correction
Calibration reference surface
Electric length compensation
Fixture compensation
Step 1. Click
Step 2. Click the
Step 3. In the
calibration/fixture compensation data. For details on the measurement points of the
calibration data, refer to Table 3-7, "Types of measurement points for calibration/fixture
compensation data," on page 60, and Table 3-8, "Cal Type box settings and measurement
points for calibration/fixture compensation data," on page 61.
Step 4. Connect the 0 S (OPEN) standard to the 7-mm terminal of the test head (see Figure 3-31 on
page 62).
Chapter 5
Impedance measurement
7-mm terminal (usual)
Required
OPEN/SHORT
compensation
on the
Cal/Comp...
Stimulus
button.
Cal Menu
box, select the desired type of measurement points for the
Cal Type
Basic Operations for Magnetic Measurement
Magnetic measurement
7-mm terminal
Not required
SHORT compensation
only
menu.

STEP 4. Calibration

133