Basic Operations for Dielectric Measurement

STEP 5. Entering Thickness of Load Standard

STEP 5. Entering Thickness of Load Standard

Enter the thickness of the PTFE load standard supplied with the 16453A test fixture by
following these steps. The thickness is printed on the surface of the case. When you use a
user-defined load standard for measuring dielectric materials, enter its thickness.
NOTE
The thickness value written on the case is a typical value. If you need to enter a highly
accurate value, measure it with a micrometer or calipers.
Step 1. Click
Step 2. Click the
Step 3. In the
standard is 0.75 mm in thickness, type
NOTE
The load standard supplied with the 16453A test fixture is made of PTFE with a relative
permittivity of 2.1. Therefore, when the E4991A is in the initial state, the value in the
Real
0.0000. If you use a user-defined load standard, change these values accordingly. For more
on the calibration kit, refer to Chapter 4, "Calibration and Compensation" in the E4991A
Operation Manual.
114
on the
Cal/Comp...
Stimulus
button.
Cal Kit Menu
box, enter the thickness of the load standard. For example, if the load
Thickness
box in the Cal Kit toolbar is set to 2.1000 and the value in the
menu.
[0] [.] [7] [5] [m] [Enter]
with the keyboard.
εr
box is set to
εr Loss
Chapter 4