Basic Operations for RF Devices Measurement
STEP 6. Fixture Compensation

Measuring SHORT Compensation Data (16197A)

Perform SHORT compensation to correct residual impedance due to the test fixture.
Step 1. Connect a short device to the test fixture. Follow the procedure given in Figure 3-50 to set
the DUT connection terminal of the test fixture to the SHORT state. When using other test
fixtures, follow the procedure described in the Operation Manual of the test fixture.
Figure 3-50
SHORT state (16197A)