Basic Operations for Magnetic Measurement
STEP 4. Calibration
Step 5. Click the
of the OPEN calibration data measurement, a √ mark appears on the left side of the
Open
NOTE
During calibration data measurement, the message "Wait-Measuring Cal Standard"
appears at the left end of the status bar at the bottom of the screen.
Step 6. Connect the 0 Ω(SHORT) standard to the 7-mm terminal of the test head (see Figure 3-33
on page 63).
Step 7. Click the
of the SHORT calibration data measurement, a √ mark appears on the left side of the
Short
Step 8. Connect the 50 Ω (LOAD) standard to the 7-mm terminal of the test head (see Figure 3-36
on page 64).
Step 9. Click the
of the LOAD calibration data measurement, a √ mark appears on the left side of the
Load
Step 10. Measure LOW-LOSS CAPACITOR calibration data, if necessary. Connect the
LOW-LOSS CAPACITOR to the 7-mm terminal of the test head (see Figure 3-38 on
page 65). Then click the
CAPACITOR calibration data. Upon completion of the LOW-LOSS CAPACITOR
calibration data measurement, a √ mark appears on the left side of the
button.
NOTE
When you measure a low loss magnetic material, measure LOW-LOSS CAPACITOR
calibration data. This allows high accuracy for high Q measurement at high frequencies.
Step 11. Click the
the measured calibration data and save it to the internal memory.
Step 12. Depending on the measurement points of the calibration/fixture compensation data
specified in the
the bottom of the screen will change, as shown in Table 4-4 on page 119.
134
button to start measuring OPEN calibration data. Upon completion
Meas Open
button.
button to start measuring SHORT calibration data. Upon completion
Meas Short
button.
button to start measuring LOAD calibration data. Upon completion
Meas Load
button.
Meas Low Loss C
button to instruct the E4991A to calculate the calibration coefficient from
Done
box, the display below the
Cal Type
button to start measuring LOW-LOSS
Meas Low Loss C
button and the status bar at
Cal Menu
Meas
Meas
Meas