STEP 6. Fixture Compensation

With the 16454A test fixture, you perform only SHORT compensation to correct residual
impedance due to the test fixture. OPEN compensation is not performed because errors due
to stray admittance are so small that they can be neglected.
Step 1. Click
Step 2. Click the
Step 3. Set the MUT connection plane of the test fixture to the SHORT state (Figure 5-7).
Figure 5-7
SHORT state (16454A)
Step 4. Click the
completion of the SHORT compensation data measurement, a √ mark appears on the left
side of the
Step 5. Click the
coefficient from the measured fixture compensation data and save it to the internal
memory.
Chapter 5
on the
Cal/Conpen...
Stimulus
button.
Comp Menu
button to start measuring SHORT compensation data. Upon
Meas Short
button.
Meas Short
button to instruct the E4991A to calculate the fixture compensation
Done
Basic Operations for Magnetic Measurement

STEP 6. Fixture Compensation

menu.
137