Basic Operations for Dielectric Measurement
STEP 10. Changing Sweep Conditions
STEP 10. Changing Sweep Conditions
When the measurement points for calibration are user-defined frequency/user-defined
power points, start measurement with "STEP 6. Calibration" on page 115 after changing
the measurement conditions. When the measurement points for calibration are fixed
frequency/fixed power points, start measurement with "STEP 9. Measuring MUT and
Analyzing Measurement Results" on page 122 after changing the measurement conditions.
NOTE
When the measurement points for calibration are fixed frequency/user-defined power
points, start measurement with "STEP 6. Calibration" on page 115 after changing the
oscillator level. When you change other sweep conditions, start measurement with "STEP
9. Measuring MUT and Analyzing Measurement Results" on page 122.
STEP 11. Measuring Other MUTs
If you measure another MUT of the same thickness as the one in the previous
measurement, start measurement with "STEP 8. Connecting MUT" on page 121. If you
measure a MUT of a different thickness, start measurement with "STEP 7. Entering
Thickness of MUT" on page 120.
Chapter 4
123