SC No.
Type
SC No.
Type
SC582-00
C
SM
Error Name/Error Condition/Major Cause/Solution
Temperature sensor defective
1. Reconnect the following connector.
Temperature Sensor – IOB connector
2. If the connector is broken, shorted, or grounded,
replace the connector.
3. Replace the temperature sensor.
IOB defective
1. Update the engine software to the latest version.
2. Replace the IOB.
Error Name/Error Condition/Major Cause/Solution
Temperature Sensor Shorted Error
The average of the output from the temperature sensor is 31 or
less for 5 seconds.
Temperature sensor defective
IOB defective
Temperature sensor defective
1. Reconnect the following connector.
Temperature Sensor – IOB connector
2. If the connector is broken, shorted, or grounded,
replace the connector.
3. Replace the temperature sensor.
IOB defective
1. Update the engine software to the latest version.
2. Replace the IOB.
6-43
Service Call 501-582
D255/D256/M281/M282