U s i n g t h e D a t a P o r t

Overview

N-600x Operator's Manual
Patient data can be output through the data port on the back of the
N-600x by connecting it to a PC or serial printer.
When connecting the N-600x to a printer or PC, verify proper
operation prior to clinical use. Both the N-600x and the printer or PC
must be connected to a grounded AC power outlet. The N-600x
protocol setting must be ASCII.
Any printer or PC connected to the monitor's data port must be
certified according to IEC Standard 950. All combinations of
equipment must be in compliance with IEC Standard 60601-1-1
systems requirements. Anyone who connects a printer or PC to the
data output port configures a medical system and is therefore
responsible for ensuring the system complies with the requirements
of system standard IEC Standard 60601-1-1 and the electromagnetic
compatibility system standard IEC Standard 60601-1-2.

Using the Data Port

111