Note: The N-395 will time out in 5 minutes when the SET and
EXIT softkeys are displayed. However, it will time out
in 10 seconds afier the SET button is pressed if there are
no further button presses.
Press the EXIT button to accept the new settings. Press EXIT
again to return to the previous menu.
Press the NEXT softkey to access the COMM softkey. The
COMM softkey provides selection of the baud rate and the
protocol of the data port.
After pressing the COMM softkey, use the ADJUST UPIDOWN
Buttons to select a baud rate of 2400, 9600, or 19200. Press
Exit to return to the SETUP submenu.
Press the SELECT softkey to highlight the protocol setting. Use
the ADJUST UPIDOWN Buttons to select the desired setting.
ASCII is used during nonnal operation and for serial printers.
GRAPH is used for graphical trend printouts when connected to
a serial printer. Select OXINET when connecting to an Oxinet"
1 1 central station network or Score software. The Bedside
Monitor Interface selections are (contained in N-395 software
versioils 1.7 and higher) AGILENT (for Agilent HP monitors),
SPACELBS (for SpaceLab monitors), MARQ (for Marquette
monitors), and DATEX (for Datex-Ohmeda AS13 monitors). Do
not use the CLINICAL setting unless instructed to do so by
Mallinckrodt's Technical Services personnel. Press Exit to
return to the SETUP submenu.
Bedside Monitor Interface
The bedside monitor interface (contained in N-395 software
versions 1.7 and higher) allows the N-395 monitor to
communicate real-time monitoring information to a "host"
bedside monitor. The purpose of bedside monitor interface is to
allow integrate
xis smart*
XL oximetry technology into the host
system for remote monitoring, trending, data storage, and other
features offered by the "liost" system.