DATA PORT PROTOCOL
Overview
Connecting to the Data Port
Baud Rate
Setting Data Port Protocol
Real-Time Display Format
Trend Data Printout (ASCII Mode)
Trend Data Printout (Graph Mode)
Nurse Call
Analog Outputs
OVERVIEW
Patient data can be obtained through the data port on the back of
the N-395 by coilnecting it to an attached PC or serial printer.
When connecting tlie N-395 to a printer or PC, verify proper
operation before clinical use. Both the N-395 and the printer or
PC must be connected to a grounded AC outlet. The N-395
protocol setting must be in the ASCII mode as described in the
"COMM" paragraph of the Start-up and Use section.
Ally printer or PC coniiected to the monitor's data port must be
certified according to IEC Standard 950. All combinations of
equipment must be in compliaiice with IEC Standard 60 1-1-1
systems requirements. Anyolie who coiinects a printer or PC to
the data output port configures a medical system and is therefore
responsible for ensuring that the system complies with the
requirements of system standard IEC Standard 60 1
-
1
-
1 and the
electromagnetic compatibility system standard IEC Standard
601-1-2.
CONNECTING TO THE DATA PORT
The N-395 data port may be connected to tlie printer or PC by
using a cable tenniilated with an AMP conilector (AMP part
number 7475 3 8- I), fermle (AMP part number 1-7475 79-2), and
compatible pins (AMP part number 66570-2). The cable should
be no more than 25 feet (7.6 meters) in length. The external ITE