Verifying The Orbital Shaking Station Configuration - Agilent Technologies AssayMAP Bravo Installation Manual

Platform
Table of Contents

Advertisement

Verifying the Orbital Shaking Station configuration

Verifying the Orbital Shaking Station configuration in a profile
AssayMAP Bravo Platform Installation Guide
IMPORTANT
If the Orbital Shaking Station is installed in the AssayMAP
Bravo Platform, use the following procedures to configure and test the
accessory in every profile.
To verify the Orbital Shaking Station configuration in a profile:
1
In Bravo Diagnostics, ensure the correct profile is selected in the Profiles tab.
2
In the Configuration tab, verify that the deck image shows a Orbital Shaking
Station icon at deck location 9.
3
To verify the settings, right- click deck location 9 in the image. In the
shortcut menu that appears, click Configure location > Accessory. The Bravo
Accessories Wizard opens.
4
In the Location for accessory list, verify the location and then click Next.
5
In the Accessory already configured page, click Yes, reconfigure existing accessory,
and then click Next.
Setting up the computer and software

Verifying the Orbital Shaking Station configuration

49

Advertisement

Table of Contents
loading

Table of Contents