Agilent Technologies AssayMAP Bravo Installation Manual page 36

Platform
Table of Contents

Advertisement

Installing the platform
Installing and connecting the Orbital Shaking Station
28
4
Insert the two M3 x 18- mm socket- head cap screws into the holes at the
front and back end of the adapter pad and use a 2.5- mm hex driver to
tighten the screws.
Tighten the screws evenly to ensure the feet on the bottom of the Orbital
Shaking Station are pressed evenly into the adapter pad, and the station
remains level.
Note: The feet of the Orbital Shaking Station are compressible, and these
screws do not reach a hard stop as a result.
CAUTION
The Orbital Shaking Station can shift slightly when tightening. Ensure that
the sides of the Orbital Shaking Station are parallel to the sides of the adapter plate. If the
Orbital Shaking Station is rotated, it could cause the Bravo gripper to collide with the
accessory.
AssayMAP Bravo Platform Installation Guide