About verifying the deck configuration in the profile

Profile
AssayMAP 1 profile
GlykoPrep- Plus
profile
AssayMAP Peptide
Sample Prep profile
AssayMAP Bravo Platform Installation Guide
Autofill Station at deck location 1. Controls pumping the liquid to and from the
96AM Wash Station. This configuration is required for all AssayMAP
applications and included in all AssayMAP profiles by default.
Heating Shaking Station at deck location 4. Controls the Peltier Thermal Station
(also known as the CPAC Ultraflat).
Note: The Heating Shaking Station configuration is required for any
heating or shaking accessory that is controlled by an STC or MTC
controller.
Orbital Shaking Station at deck location 9. Controls the Orbital Shaking Station.
The following table describes the requirements for each profile.
Verification tasks
By default, the AssayMAP 1 profile
includes only the Autofill Station (96AM
Wash Station) as an accessory
configuration.
All other deck locations are configured as
standard platepads.
If the Peltier Thermal Station and Orbital
Shaking Station are installed on the
AssayMAP Bravo deck, you configure the
corresponding deck locations for these
accessories in the AssayMAP 1 profile.
By default, the GlykoPrep- Plus profile
includes the following configurations:
Autofill Station (96AM Wash Station)
Heating Shaking Station for the Peltier
Thermal Station (location 4)
Orbital Shaking Station (location 9)
If these accessories are not installed on the
AssayMAP Bravo deck, you configure the
corresponding deck locations for a
standard platepad.
By default, the AssayMAP Peptide Sample
Prep profile includes the following
configurations:
Autofill Station (96AM Wash Station)
Heating Shaking Station for the Peltier
Thermal Station (location 4)
Orbital Shaking Station (location 9)
If these accessories are not installed on the
AssayMAP Bravo deck, you configure the
corresponding deck locations for a
standard platepad.
Setting up the computer and software

About verifying the deck configuration in the profile

See...
"Verifying the Autofill Station
configuration (96AM Wash
Station)" on page 45
"Verifying the Autofill
Station configuration (96AM
Wash Station)" on page 45
"Verifying the Peltier
Thermal Station
configuration" on page 47
"Verifying the Orbital
Shaking Station
configuration" on page 49
"Verifying the Autofill
Station configuration (96AM
Wash Station)" on page 45
"Verifying the Peltier
Thermal Station
configuration" on page 47
"Verifying the Orbital
Shaking Station
configuration" on page 49
43