Agilent Technologies AssayMAP Bravo Installation Manual page 102

Platform
Table of Contents

Advertisement

Verifying the installation and setup
Performing a mock run
94
4
In the Labware Table area, use the default settings for the labware selections.
5
Click Run Affinity Purification to start the run.
The protocol run starts and checks for conflicting step selections. The
protocol with the 500- µL/min flow rate should take less than 15 minutes
to complete.
6
During the mock run, observe the automation movements of the Bravo
96AM Head to assess whether any problems occur. Make a note of any
problems. See the following table for the sequence of movements.
To monitor the progress of the run, check the Status box.
To pause the run, click Pause. The task currently in progress finishes
before the protocol pauses.
WARNING
To stop a run in an emergency, use the hardware Emergency Stop button.
7
After the run finishes, remove all labware from the AssayMAP Bravo deck.
AssayMAP Bravo Platform Installation Guide