Setting up the computer and software
Verifying the Peltier Thermal Station configuration
48
If the accessory is not yet configured at this location, in the Choose
Accessory page, choose the Heating Shaking Station.
Note: Earlier versions of Bravo Diagnostics use the name Inheco STC/
MTC Controller for this accessory.
6
Verify the following settings in the Properties table, and then click Next.
Properties
MTCSTC Id
Type
Slot Id
Note: The Inheco STC Controller ships with an MTCSTC Id setting of 0.
However, this setting may have been changed on demo units for use
with multiple devices. Use the DIP switches on the back of the STC
Controller to determine what the actual Id setting is and make changes
if necessary. See the manufacturer's user documentation for more
information.
7
Click Finish. The Bravo Accessories Wizard closes.
8
When the accessory configuration message appears, click Yes to move the
teachpoint to a safe height.
9
If you made any changes to the configuration settings, click the Profiles tab,
and then click Update this profile to save the changes.
Setting
0
STC
1
AssayMAP Bravo Platform Installation Guide