Agilent Technologies AssayMAP Bravo Installation Manual page 23

Platform
Table of Contents

Advertisement

Figure Tubing configuration
4
1
Item
Component
1
Source bottle
2
Waste bottle
3
Tubing, 8- mm
(5/16- in)
4
Tubing, 6.4- mm
(1/4- in)
5
Connector,
union
6
Quick-
disconnect
fitting
7
Inline pump
filter
8
3- way
connector
AssayMAP Bravo Platform Installation Guide
6
4
5
3
4
6
2
Description
Supplies the fill liquid for the wash station.
Collects the waste liquid that is pumped from the reservoir.
The length of tubing in the lower pump head, which is used for emptying
the wash station. The larger diameter tubing ensures that the fluid is
removed efficiently from the wash station
The fill- line tubing from the source bottle to the wash station, and
empty- line tubing connections on either side of the 5/16- in diameter
tubing.
Note: The 5/16- in diameter tubing is too large for the quick- disconnect
fittings and the wash station ports. So, you use a union connector (5) to
splice the 5/16- in tubing to the 1/4- in tubing.
Joins the two sizes of tubing.
Enables easy removal of the wash station and quick replacement of the
tubing in the pump heads. The quick- disconnect fittings include an
automatic- close valve. Fluid can flow only if a positive connection is
made.
Removes the particulates that can clog the chimneys in the wash station.
Enables one tube to branch into two tubes at the input and output ports
on the wash station.
Installing the 96AM Wash Station and Pump Module
7
6
4
8
3
5
4
6
4
8
Installing the platform
15