Initializing The Profile That You Are Verifying - Agilent Technologies AssayMAP Bravo Installation Manual

Platform
Table of Contents

Advertisement

Initializing the profile that you are verifying

Before you start
Initializing the profile in Bravo Diagnostics
AssayMAP Bravo Platform Installation Guide
Initializing the profile establishes communication with the AssayMAP Bravo
Platform and accessories. You will initialize a profile to test the accessory
settings and verify the gripper adjustment for that profile.
WARNING
When you initialize the Bravo Platform, the head and tie bar can move. To
prevent potential injury, keep clear of the device while it is in motion.
To initialize the profile:
1
In the Profiles tab, ensure that the correct profile is selected, and then click
Initialize this profile to start initializing the selected profile.
2
If the There appears to be a plate present error message appears, click Ignore and
Continue to continue the homing process.
3
When the Verify that it is safe to home the W-axis message appears, click Retry to
continue homing the w- axis.
Setting up the computer and software

Initializing the profile that you are verifying

51