Manually updating teachpoints in each profile
Adjusting the teachpoint for the red PCR plate insert
Adjusting the teachpoint for the PCR plate insert
104
To adjust the teachpoint for the PCR plate insert:
1
In the Jog/Teach tab, click Location 4.
2
Click Move to move the head to the previously defined teachpoint for
location 4.
3
Using the Jog Axes controls, jog the head up (- Z) by exactly 4.4 mm.
For example, if the value for the teachpoint's Z- axis coordinate from the
fine tuning was 104.00 mm, the new value after adjusting by 4.4 mm will
be 99.60 mm.
4
Click Teach to set the new teachpoint.
5
To save the new teachpoint for this location, click the Profiles tab, and then
click Update this profile.