Installing the Bravo risers

Before you start
Procedure
AssayMAP Bravo Platform Installation Guide
WARNING
The Bravo device weighs approximately 52.1 kg (114.9 lb). Attempting to
move the Bravo device without assistance could cause personal injury. Request assistance
and use proper lifting techniques when moving the Bravo device.
CAUTION
Tugging on the tie bar or using it to lift the Bravo Platform can damage the
device. Do not grab the tie bar when lifting the Bravo Platform.
00233
Bravo
Proper lift
IMPORTANT
The following procedure is intended only for fully qualified and
trained Agilent service engineers.
Ensure that you have the following:
Two people to lift the Bravo Platform
Foam pad (3- inch thick) or similar cushioning material to use when you
tilt the Bravo Platform onto its back plate
Note: EPE foam is preferred due to its firmness.
Risers
Hex drivers
CAUTION
When tilting the Bravo Platform backwards to rest on its back plate, the head
mount can slide back rapidly along the y-axis and collide with the back plate. To prevent
potential equipment damage, ensure that the y-axis shipping stop remains installed on the
head mount until after you install the risers. If the shipping stop has already been removed
during unpacking, ensure that the head mount is resting against the Bravo back plate before
tilting the Bravo Platform backwards.
To install the risers:
1
Place the foam pad on the work table behind the Bravo Platform.
00233
00234
Installing the platform

Installing the Bravo risers

00234
Bravo
Improper lift
5