Running the Syringe Test utility

Required reagents, labware, and equipment
Item
Required items
1
Syringe Test Kit
2
96- well microplate,
(2 per test)
3
Open reservoir,
(2 per test)
4
Centrifuge
5
Microplate reader
Workflow
AssayMAP Bravo Platform Installation Guide
You use the AssayMAP Syringe Test Kit to evaluate all the syringes in the head
concurrently while the Bravo 96AM Head is mounted on the AssayMAP Bravo
Platform.
Description
Contains materials and reagents to
perform 20 tests of the 96 syringes
in the Bravo 96AM Head.
Flat bottom, clear polystyrene
microplate for test sample collection.
Open reservoir to hold water and
diluted test solution.
Centrifuge with a microplate rotor
capable of achieving 300 x g
Spectrophotometer capable of
detecting absorbance at 425 nm in a
96- well microplate and generating a
text (TXT) file of the data.
To run the Syringe Test utility:
1
Open the Utility Library, locate the Syringe Test utility, and then click Utility.
2
Click Instructions.
3
Follow the instructions to run the Syringe Test.
4
If required, replace defective syringes, and then rerun the Syringe Test to
confirm all the syringes are functioning properly.
Verifying the installation and setup

Running the Syringe Test utility

Part number
G5496- 60050
Greiner 655101
Nunc 269620
Corning 9017
Agilent G5498B #049
Agilent 201254- 100
Axygen RES- SW96- LP
Not applicable
NA
97