Verifying the Autofill Station configuration (96AM Wash Station)

Verifying the Autofill Station configuration in a profile
AssayMAP Bravo Platform Installation Guide
IMPORTANT
Use the following procedure to verify the Autofill Station
configuration settings in every AssayMAP profile.
To verify the Autofill Station configuration in a profile
1
In Bravo Diagnostics, ensure the correct profile is selected in the Profiles tab.
2
In the Configuration tab, verify that the deck image shows a Pump Module at
deck location 1.
Note: The AssayMAP 1 profile has no other accessories configured by
default.
3
To verify the settings, right- click deck location 1 in the image. In the
shortcut menu that appears, click Configure location > Accessory. The Bravo
Accessories Wizard opens.
4
In the Location for accessory list, verify the location and then click Next.
5
In the Accessory already configured page, click Yes, reconfigure existing accessory,
and then click Next.

Verifying the Autofill Station configuration (96AM Wash Station)