AssayMAP Bravo Platform Installation Guide
To test the emergency stop feature:
1
Open the Utility Library, locate the Cartridge Transfer utility, and then click
Utility.
2
In the A. Application Settings area, specify the following settings for the run:
Setting
First Column in Source
Cartridge Rack
Number of Columns to Transfer 12
First Column in Destination
Cartridge Rack
3
Click Run Cartridge Transfer to start the run.
4
As soon as the Bravo head starts to move away from its position above the
wash station and towards deck location 6, press the red button on the
pendant.
The Bravo head stops moving and an error message appears.