Setting up the computer and software
About verifying the deck configuration in the profile
Profile
Single Plate
InSolution Digestion
profile
AssayMAP
Normalization profile
44
Verification tasks
By default, the Single Plate InSolution
Digestion profile includes the following
configurations:
Autofill Station (96AM Wash Station)
Heating Shaking Station for the Peltier
Thermal Station (location 4)
Orbital Shaking Station (location 9)
If these accessories are not installed on the
AssayMAP Bravo deck, you configure the
corresponding deck locations for a
standard platepad.
In addition to the physical accessories, this
profile includes a Lid Hotel configuration
at deck locations 3 and 7. Even though
standard platepads are installed at these
deck locations, the Lid Hotel configuration
enables the locations for Delidding and
Relidding tasks.
By default, the AssayMAP Normalization
profile includes the following
configurations:
Autofill Station (96AM Wash Station)
Heating Shaking Station for the Peltier
Thermal Station (location 4)
If Peltier Thermal Station is not installed
on the AssayMAP Bravo deck, you
configure the corresponding deck location
for a standard platepad.
The AssayMAP Normalization profile also
includes accessory teachpoints at deck
locations 2 and 6 for the Bravo Plate Riser
accessory, which must be verified and
adjusted if necessary. The Bravo Plate
Riser has no configuration settings.
See...
"Verifying the Autofill
Station configuration (96AM
Wash Station)" on page 45
"Verifying the Peltier
Thermal Station
configuration" on page 47
"Verifying the Orbital
Shaking Station
configuration" on page 49
"Verifying the Autofill
Station configuration (96AM
Wash Station)" on page 45
"Verifying the Peltier
Thermal Station
configuration" on page 47
AssayMAP Bravo Platform Installation Guide