Setting the teachpoints
Teaching part 1: Setting teachpoints in AssayMAP 1 profile
60
2
In the Bravo Diagnostics dialog box, click the Jog/Teach tab.
3
In the image of the Bravo deck in the upper left corner, click Location 2.
4
Carefully, use the Jog Axes controls to adjust the position of the pipette tip
above the crosshairs on platepad 2. Use incrementally smaller jog
increments as you jog the pipette tip down closer to the crosshairs.
As the pipette tip moves down along the z- axis towards the crosshairs, the
following error message appears. Visually inspect the clearance remaining
between the tip and the crosshairs. If the tip is not yet at the crosshairs,
adjust the jog increment as required, and click Ignore to continue jogging
the head down.
CAUTION
Before you click Ignore in the error message, verify that enough space
remains to continue jogging the head down. Jogging the head down too far will result in
a collision and cause equipment damage.
Note: This error message appears every time you jog in the z- axis to a
point that is lower than the default z- axis teachpoint (60.00 mm).
5
When the pipette tip is 1–2 mm above the crosshairs, do the following:
AssayMAP Bravo Platform Installation Guide