Testing the Pump Module
AssayMAP Bravo Platform Installation Guide
Use the following procedure to test the communication with the Pump Module.
In a subsequent verification procedure, you will run the Startup utility, which
will test the functioning of the Pump Module and wash station using liquid.
To test the communication with the Pump Module:
1
Ensure that the Pump Module is on.
2
Lift the flip- top cover on each pump head so that it is fully open.
Note: While the pump heads are open, no liquid will pump in the
subsequent steps.
00306
3
In the Configuration tab, click the Autofill Station icon, and then click Diagnose
accessory. The Bravo Accessory Diagnostics dialog box opens.
Setting up the computer and software
Testing the accessories
53